# Spectrum Requirements for Advanced Medical Technologies
> **Federal Communications Commission** · Final rule. · Published 2010-08-26 · Effective 2010-09-27 · 75 FR 52472
## Document
- **Document number:** 2010-21011
- **Category:** spectrum-allocation
- **Federal Register citation:** 75 FR 52472
- **CFR reference:** 47 CFR 95
- **Publication date:** 2010-08-26
- **Effective date:** 2010-09-27
- **FCC docket:** ET Docket No. 06-135
## Abstract

This document addresses a petition for reconsideration (petition) filed by Medtronic, Inc. (Medtronic) regarding rules for the Medical Device Radiocommunication (MedRadio) service. The Commission grants reconsideration to the extent of amending the MedRadio rules to permit the submission of average power transmitter measurements, and making editorial corrections or clarifications to several provisions concerning the frequency monitoring criteria and permissible communications for "listen-before-talk" (LBT) and non-LBT devices. The Commission denies reconsideration in all other respects and otherwise affirms certain provisions of the MedRadio rules questioned by Medtronic.

## Source
- [Federal Register document](https://www.federalregister.gov/documents/2010/08/26/2010-21011/spectrum-requirements-for-advanced-medical-technologies)
---
*AI Analytics · CC0 1.0*