{"url_path":"/fda/device/recalls/Z-1976-2026","section_key":"summary","section_title":"DEVICE Recall: AVID Medical, Inc. — Halyard NEURO IR PERIPHERAL (PS 64405) kit. Model Number: UINR44-01.","topic":"fda","document":{"doc_type":"device-recall","doc_date":"2026-03-13","source_url":"https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1976-2026%22","accession_number":null,"cik":null,"ticker":null,"issuer_name":null,"edgar_url":null,"primary_entity_key":null,"primary_entity_name":"AVID Medical, Inc."},"word_count":93,"has_tables":false,"body_markdown":"# FDA DEVICE Recall: Z-1976-2026\n\n**Recalling firm:** AVID Medical, Inc.\n**Classification:** Class I\n**Status:** Ongoing\n**Type:** Voluntary: Firm initiated\n**Initiated:** 2026-03-13\n**Report date:** 2026-05-13\n\n## Product\nHalyard NEURO IR PERIPHERAL (PS 64405) kit. Model Number: UINR44-01.\n\n## Reason\nPotential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.\n\n## Affected lots\nModel Number: UINR44-01. UDI (Kit Lot Number): (01)20809160484439(17)280201(10)340135(30)004(241)UINR44-01 (340135), (01)20809160484439(17)280201(10)340136(30)004(241)UINR44-01 (340136), (01)20809160454439(17)280201(10)340352(30)004(241)UINR44-01 (340352), (01)20809160484439(17)280201(10)349076(30)004(241)UINR44-01 (349076).\n\n## Distribution\nUS distribution to FL, IL, MO, MS, NC, NE, TX."}