{"url_path":"/fda/device/recalls/Z-2648-2026","section_key":"summary","section_title":"DEVICE Recall: Siemens Medical Solutions USA, Inc — ARTIS icono ceiling. Model Number: 11328100.","topic":"fda","document":{"doc_type":"device-recall","doc_date":"2026-06-08","source_url":"https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2648-2026%22","accession_number":null,"cik":null,"ticker":null,"issuer_name":null,"edgar_url":null,"primary_entity_key":null,"primary_entity_name":"Siemens Medical Solutions USA, Inc"},"word_count":94,"has_tables":false,"body_markdown":"# FDA DEVICE Recall: Z-2648-2026\n\n**Recalling firm:** Siemens Medical Solutions USA, Inc\n**Classification:** Class II\n**Status:** Ongoing\n**Type:** Voluntary: Firm initiated\n**Initiated:** 2026-06-08\n**Report date:** 2026-07-08\n\n## Product\nARTIS icono ceiling. Model Number: 11328100.\n\n## Reason\nPotential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.\n\n## Affected lots\nModel Number: 11328100. UDI-DI: 4056869295923.\n\n## Distribution\nUS Nationwide distribution in the states of CA, GA, IA, IN, LA, MO, NC, NY, PA, SC."}