# High-order Cavities for Enabling High-field MAS-DNP-NMR with Low-cost THz Sources

> **NIH NIH R43** · DOTY SCIENTIFIC, INC. · 2021 · $300,001

## Abstract

High-order Cavities for Enabling High-field MAS-DNP-NMR with Low-cost THz Sources
Abstract
Recently the first atomic-resolution structures of the Aβ40 and Aβ42 amyloid fibrils that play an important role in
Alzheimer’s Disease (AD) were solved by a combination of methods that relied critically on solid-state NMR
(ssNMR). Key to that structure determination was a technique denoted as dynamic nuclear polarization (DNP)
with magic angle spinning (MAS). While Cryo EM, scanning tunneling electron microscopy (STEM) and other
methods provided much useful information and low-resolution structures, recent advances in MAS-NMR 
methods provided essential restraints and additional crucial information, including sidechain dynamics important in
protein functions and in understanding of myriad mechanisms of their action. Hence, developing 
transformational advances for ssNMR is crucial for both structural biology and biomedical research in general, for 
progress in addressing Alzheimer’s Disease, novel viruses, cancer, arthritis, and other inflammasome-related 
diseases, particularly for providing regio-specific drug binding information enabling detailing of the mechanism of
action for effective drugs.
High-field (>11 T) MAS-DNP systems thus far have all required the high microwave power only available from
gyrotrons. Because of their narrow bandwidth, they then require an NMR magnet with superconducting sweep
coils and thus directly and indirectly add $1.5-3M to the system cost. Moreover, the DNP methods are limited
by the poor frequency agility of gyrotrons, and many crowded NMR laboratories simply do not have space to
accommodate a gyrotron or the funds to acquire one.
While a recent advance in a commercially available 1.3-mm MAS-DNP probe showed improved microwave
efficiency at 9.4 T (400 MHz/263 GHz), our detailed simulations show that still only 2-9% of the incident micro-
wave power is dissipated within the lossy sample. This suggests there is room for another order of magnitude
improvement in microwave efficiency, and our preliminary simulations of a novel high-order microwave cavity
compatible with a novel MAS spinner design further support this goal.
This proposed Phase I SBIR will carry out in silico evaluations of novel fast-MAS spinner designs with integral
high-order microwave cavities that promise order-of-magnitude increases in microwave efficiency at 
frequencies in the 195-528 GHz range (as needed for NMR at 7-18.8 T). Experimental evaluations of novel cavity-
MAS-DNP modules will be carried out using a solid-state source on the bench and in a triple-resonance
500MHz/328GHz 1.3-mm MAS-DNP probe at 11.7 T. It is expected that this advance, in combination with 
several other technological advances being pursued in other projects, will eventually enable DNP to be added to
existing ssNMR high-field systems without the requirement of either a specialized magnet or a gyrotron.

## Key facts

- **NIH application ID:** 10147206
- **Project number:** 1R43GM140650-01
- **Recipient organization:** DOTY SCIENTIFIC, INC.
- **Principal Investigator:** Francis DAVID Doty
- **Activity code:** R43 (R01, R21, SBIR, etc.)
- **Funding institute:** NIH
- **Fiscal year:** 2021
- **Award amount:** $300,001
- **Award type:** 1
- **Project period:** 2021-02-01 → 2022-10-31

## Primary source

NIH RePORTER: https://reporter.nih.gov/project-details/10147206

## Citation

> US National Institutes of Health, RePORTER application 10147206, High-order Cavities for Enabling High-field MAS-DNP-NMR with Low-cost THz Sources (1R43GM140650-01). Retrieved via AI Analytics 2026-05-25 from https://api.ai-analytics.org/grant/nih/10147206. Licensed CC0.

---

*[NIH grants dataset](/datasets/nih-grants) · CC0 1.0*