# Contamination resistant interconnection for reducing peritoneal dialysis - associated peritonitis

> **NIH NIH R41** · CEROVATIONS, LLC · 2020 · $314,354

## Abstract

Abstract
Significance: Peritonitis remains a significant challenge in the management of end stage renal disease
(ESRD) for patients on peritoneal dialysis (PD) and contributes significantly to morbidity and mortality. More
than half of these infections result from bacteria originating on skin or mucous membranes, suggesting patient
or caregiver contamination as the source of infections. Furthermore, reviewed literature, results from a 6 center
survey and 57 interviews with PD providers all suggest contamination during connection for therapy as a
common root cause of peritonitis. Patient factors are important in contamination control as existing strategies
focus on patient training and depend on patient compliance to be successful. However, fewer than 50% of
patients remain compliant to these trainings after 6 months of starting PD. Innovation: The proposed product
is a low cost, low risk and patient friendly connection interface that can prevent contamination without
dependence on patient compliance. Connection for therapy in this design occurs within an enclosure that
shields from touch and aerosol contamination. Scientific Premise: This project is based on the hypothesis
that eliminating touch and airborne contamination will reduce peritonitis by up to 50%, the current infection rate
by organisms colonizing the skin or mucous membranes. Preliminary Data: Applicant has performed bench
experiments demonstrating proof of concept for the contamination resistant interconnection and the in vitro
microbiology test method. STTR Phase I Project Objective: Demonstrate feasibility of a contamination
resistant PD interconnection that reduces bacterial transfer to dialysis fluid path. Aim 1: Freeze design for
contamination resistant interconnection. Contamination resistant interconnection design will be finalized to
meet human use requirements based on Aim 1 PD design inputs from clinical experts and patient usability
study. Contamination resistant interconnection design will be built, assessed, refined and finalized for Aim 2
testing. The bench testing will follow test protocols and standards utilized in 510K summaries of similar,
predicate interconnection devices. Aim 1 will focus on meeting design requirements for leak resistance, fluid
flow, and usability. Milestone: Bench testing will demonstrate the potential for meeting Aim 2 success criteria.
Aim 2: Contamination resistant interconnection feasibility testing. The objective is to demonstrate
feasibility to support STTR Phase II further research, which will focus on contamination resistant
interconnection design verification testing (DVT) for FDA submissions and clinical studies. The Aim 1 design
will be evaluated with bench tests (n = 10 devices per test) and a simulated touch and aerosol contamination
protocol using 2 commonly implicated bacterial strains. The Aim 1 final design will be compared for
contamination to an FDA cleared and marketed connection interface in the microbiology testing (n = 20...

## Key facts

- **NIH application ID:** 10149887
- **Project number:** 1R41DK126586-01A1
- **Recipient organization:** CEROVATIONS, LLC
- **Principal Investigator:** Tom Viker
- **Activity code:** R41 (R01, R21, SBIR, etc.)
- **Funding institute:** NIH
- **Fiscal year:** 2020
- **Award amount:** $314,354
- **Award type:** 1
- **Project period:** 2020-09-17 → 2022-08-31

## Primary source

NIH RePORTER: https://reporter.nih.gov/project-details/10149887

## Citation

> US National Institutes of Health, RePORTER application 10149887, Contamination resistant interconnection for reducing peritoneal dialysis - associated peritonitis (1R41DK126586-01A1). Retrieved via AI Analytics 2026-05-23 from https://api.ai-analytics.org/grant/nih/10149887. Licensed CC0.

---

*[NIH grants dataset](/datasets/nih-grants) · CC0 1.0*