fda-device · Food and Drug Administration · Rule · Published 2002-07-17 · Effective 2002-10-15 · 67 FR 46851
Document
Document number
02-17957
Federal Register citation
67 FR 46851
CFR reference
21 CFR 868
Type
Rule
Action
Final rule.
Category
fda-device
Sub-agency
Food and Drug Administration
Publication date
2002-07-17
Effective date
2002-10-15
HHS docket
Docket No. 00N-1457
Abstract
The Food and Drug Administration (FDA) is issuing a final rule to create a separate classification for the apnea monitor. The device currently is included in the generic type of device called breathing frequency monitors. The apnea monitor will remain in class II, but will be subject to a special control. The special control is an FDA guidance document that identifies minimum performance, testing, and labeling recommendations for the device. Following the effective date of this final classification rule, any firm submitting a 510(k) premarket notification for a "new" apnea monitor will need to address the issues covered in the special control guidance. However, the firm need only show that its device meets the recommendations of the guidance or in some other way provides equivalent assurances of safety and effectiveness. Elsewhere in this issue of the Federal Register, FDA is announcing the availability of the guidance document that will serve as the special control. FDA is taking these actions because it believes that they are necessary to provide reasonable assurance of the safety and effectiveness of the apnea monitor.