Medical Devices; Hematology and Pathology Devices; Reclassification of Automated Blood Cell Separator Device Operating by Filtration Principle from Class III to Class II
fda-device · Food and Drug Administration · Rule · Published 2003-02-28 · Effective 2003-03-31 · 68 FR 9530
Document
Document number
03-4690
Federal Register citation
68 FR 9530
CFR reference
21 CFR 864
Type
Rule
Action
Final rule.
Category
fda-device
Sub-agency
Food and Drug Administration
Publication date
2003-02-28
Effective date
2003-03-31
HHS docket
Docket No. 96P-0484
Abstract
The Food and Drug Administration (FDA) is reclassifying the automated blood cell separator (ABCS) device operating by filtration principle, intended for routine collection of blood and blood components, from class III to class II (special controls). The special control requirement for this device is an annual report with emphasis on adverse reactions to be filed by the manufacturer for a minimum of 3 years. The agency is taking this action in response to a petition submitted under the Federal Food, Drug, and Cosmetic Act (the act) as amended by the Medical Device Amendments of 1976 (the 1976 amendments), the Safe Medical Devices Act of 1990 (the SMDA), and the Food and Drug Administration Modernization Act of 1997 (FDAMA). The agency is reclassifying the automated blood cell separator devices operating by filtration principle into class II (special controls) because special controls, in addition to general controls, are capable of providing a reasonable assurance of safety and effectiveness of the device.