← HHS Federal Register rules

Effective Date of Requirement for Premarket Approval for Automated External Defibrillator Systems; Republication

fda-device · Food and Drug Administration · Rule · Published 2015-02-03 · Effective 2015-02-03 · 80 FR 5674

Document

Document number
2015-02049
Federal Register citation
80 FR 5674
CFR reference
21 CFR 870
Type
Rule
Action
Final order; republication.
Category
fda-device
Sub-agency
Food and Drug Administration
Publication date
2015-02-03
Effective date
2015-02-03
HHS docket
Docket No. FDA-2013-N-0234

Abstract

The Food and Drug Administration (FDA or the Agency) is republishing in its entirety a final order entitled "Effective Date of Requirement for Premarket Approval for Automated External Defibrillator" that published in the Federal Register on January 29, 2015 (80 FR 4783). FDA is republishing to correct an inadvertent omission of a comment regarding adverse tissue reaction as a risk to health and the Agency's response to that comment. The final order requires the filing of premarket approval applications (PMA) for automated external defibrillator (AED) systems, which consist of an AED and those AED accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., pad electrodes, batteries, adapters, and hardware keys for pediatric use).