Medical Devices; Exemption From Premarket Notification: Class II Devices; Surgical Apparel
fda-device · Food and Drug Administration · Rule · Published 2018-05-17 · Effective 2018-05-17 · 83 FR 22846
Document
Document number
2018-10563
Federal Register citation
83 FR 22846
CFR reference
21 CFR 878
Type
Rule
Action
Final order.
Category
fda-device
Sub-agency
Food and Drug Administration
Publication date
2018-05-17
Effective date
2018-05-17
HHS docket
Docket No. FDA-2017-N-4919
Abstract
The Food and Drug Administration (FDA or Agency) is publishing this final order to exempt certain surgical apparel from premarket notification requirements, subject to conditions and limitations. FDA is limiting the exemption to single-use, disposable respiratory protective devices (RPD) used in a healthcare setting and worn by healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. These devices, commonly referred to as N95 filtering facepiece respirators (FFRs) and surgical N95 respirators (herein collectively referred to as N95s) are currently regulated by FDA under product code MSH. This exemption will decrease regulatory burden on the medical device industry and will eliminate private costs and expenditures required to comply with certain Federal regulations. All other class II devices classified under FDA's surgical apparel classification regulation continue to be subject to premarket notification requirements. FDA is also amending the codified language for the surgical apparel devices classification regulation to reflect this final determination.