FDA recall Z-0013-2018

Handicare Usa Inc · Class II · device

Product

PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered

Reason for recall

During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.

Distribution

Us and Canada

Key facts

Status
Terminated
Initiation date
2017-09-08
Report date
2017-10-18
Termination date
2018-12-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Maryland Heights, MO, United States