FDA recall Z-0015-2018

Philips Electronics North America Corporation · Class II · device

Product

Philips Healthcare Brilliance 64 System, X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason for recall

During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.

Distribution

USA (nationwide) Distribution

Key facts

Status
Terminated
Initiation date
2017-01-18
Report date
2017-10-25
Termination date
2020-02-26
Voluntary/Mandated
FDA Mandated
Location
Andover, MA, United States

Primary source