FDA recall Z-0016-2026

Getinge Disinfection Ab · Class II · device

Product

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Reason for recall

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

Distribution

Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2025-08-29
Report date
2025-10-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Vaxjo, N/A, Sweden

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0016-2026