# FDA recall Z-0019-2025

> **Outset Medical, Inc.** · Class II · device recall initiated 2024-08-01.

## Product

Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility.

## Reason for recall

Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.

## Distribution

US: CA, NY, IL, MO, KY, AZ, PA, FL, GA, AK, NM, VA, NC, SC, OK, AL, MA, OH, MI, WA, ID, TX, TN, WY, IA, LA, NE, IN, AR, NV, WV, CO, NJ, MD, KS, UT, CT, MN, MS, WI, HI, DE, VT, MT, OR, ND.  OUS: UAE

## Key facts

- **Recall number:** Z-0019-2025
- **Recalling firm:** Outset Medical, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-08-01
- **Report date:** 2024-10-09

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** San Jose, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0019-2025

## Citation

> AI Analytics. FDA recall Z-0019-2025. Retrieved 2026-07-20 from https://api.ai-analytics.org/recall/Z-0019-2025. Source: US FDA. Licensed CC0.

---