FDA recall Z-0021-2018

OriGen Biomedical, Inc. · Class I · device

Product

The OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of the device. A thin sheath is used to cover and secure the wire layer and is heat shrunk into a permanent position.

Reason for recall

Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss).

Distribution

US Nationwide and the countries of Austria, Belgium, Chile, Czech Republic, Italy, Kingdom of Saudi Arabia, Netherlands, Poland, Sweden, Thailand, Colombia, India, South Africa, United Arab Emirates.

Key facts

Status
Terminated
Initiation date
2017-08-02
Report date
2017-11-01
Termination date
2024-02-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Austin, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0021-2018