FDA recall Z-0022-2018

St. Jude Medical, Inc. · Class II · device

Product

Proclaim 5 Elite Implantable Pulse Generator and Proclaim 7 Elite Implantable Pulse Generator Models 3660, 3661, 3662, 3663, 3665, 3667

Reason for recall

Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.

Distribution

Worldwide Distribution - US (nationwide) (foreign consignee list will be provide by the firm at a later time)

Key facts

Status
Terminated
Initiation date
2017-09-12
Report date
2017-10-25
Termination date
2020-04-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Plano, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0022-2018