FDA recall Z-0022-2025

Qualigen Inc · Class II · device

Product

FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

Reason for recall

Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.

Distribution

US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland

Key facts

Status
Ongoing
Initiation date
2024-08-26
Report date
2024-10-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Carlsbad, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0022-2025