FDA recall Z-0025-2025

Beckman Coulter, Inc. · Class II · device

Product

Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Reason for recall

When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians.

Distribution

Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.

Key facts

Status
Ongoing
Initiation date
2024-04-18
Report date
2024-10-16
Voluntary/Mandated
Voluntary: Firm initiated