# FDA recall Z-0026-2025

> **Becton Dickinson & Co.** · Class II · device recall initiated 2024-09-12.

## Product

BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30-  IVD standalone laboratory software solution providing data management and workflow management functionality across locations where clinical diagnostic activities take place.  Catalog Number: 444150

## Reason for recall

Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy

## Distribution

Nationwide Foreign: Australia, New Zealand, Singapore, Hong Kong, Taiwan, Japan, Austria, Belgium, Denmark, France, Germany, Italy, Luxembourg, Netherlands, Norway, Spain, Sweden, Switzerland, United Kingdom, Russia, Saudi Arabia, South Africa, Canada,  Mexico, China

## Key facts

- **Recall number:** Z-0026-2025
- **Recalling firm:** Becton Dickinson & Co.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-09-12
- **Report date:** 2024-10-16

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Sparks, MD, United States