FDA recall Z-0028-2022

Johnson & Johnson Surgical Vision Inc · Class II · device

Product

TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the capsular bag. The TECNIS Simplicity Delivery System is designed as a single-use disposable preloaded modular cartridge that functions as both the primary packaging and as a sterile, disposable insertion system for delivering the IOL into the eye during cataract surgery.

Reason for recall

Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin specification limit. Product was released in error based on the results of additional testing performed during the investigation into release testing nonconformance.

Distribution

U.S. Nationwide distribution in the state of IL.

Key facts

Status
Terminated
Initiation date
2021-08-06
Report date
2021-10-13
Termination date
2023-03-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Santa Ana, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0028-2022