FDA recall Z-0030-2020

Baxter Healthcare Corporation · Class II · device

Product

L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide.

Reason for recall

Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.

Distribution

U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution.

Key facts

Status
Terminated
Initiation date
2018-10-10
Report date
2019-10-09
Termination date
2020-07-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States