## Product

Citrine Triple Quad Mass Spectrometer, Part Number 5063684

## Reason for recall

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

## Distribution

US Nationwide.

## Key facts

- **Recall number:** Z-0040-2024
- **Recalling firm:** AB Sciex, LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-08-30
- **Report date:** 2023-10-11

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Framingham, MA, United States