FDA recall Z-0041-2024

Blue Belt Technologies, Inc · Class II · device

Product

Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

Reason for recall

CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2023-09-07
Report date
2023-10-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Pittsburgh, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0041-2024