FDA recall Z-0044-2018

Bard Peripheral Vascular Inc · Class II · device

Product

BARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm; MC1816, 18 g x 16 cm; MC1820, 18 g x 20 cm; MC1825, 18 g x 25 cm. The core needle biopsy device is intended for use in obtaining biopsies from soft tissues.

Reason for recall

During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and Self-Activation.

Distribution

US, Australia, Canada, Europe, Middle East, Africa, Latin America, Asia Pacific

Key facts

Status
Terminated
Initiation date
2017-09-22
Report date
2017-11-01
Termination date
2018-09-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tempe, AZ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0044-2018