FDA recall Z-0053-2025

MEDLINE INDUSTRIES, LP - Northfield · Class II · device

Product

Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket

Reason for recall

CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits.

Distribution

US Nationwide distribution in the state of CA.

Key facts

Status
Ongoing
Initiation date
2024-08-28
Report date
2024-10-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States