FDA recall Z-0055-2018

Biosense Webster, Inc. · Class II · device

Product

THERMOCOOL SMARTTOUCH SF Bi-Directional Product Codes D-1348-01-S, D-1348-04-S, D-1348-05-S

Reason for recall

Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.

Distribution

US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China

Key facts

Status
Terminated
Initiation date
2017-09-06
Report date
2017-11-08
Termination date
2019-11-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irwindale, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0055-2018