FDA recall Z-0058-2025

Merit Medical Systems, Inc. · Class II · device

Product

Monarch Inflation Device MAP 112 Access-9 Large Bore Hemostasis Value Metal Insertion Tool Torque DeviceREF IN2230 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure the pressure within the balloon during the procedure. It is also used to inject fluids into the body and to monitor the pressure of that fluid.

Reason for recall

Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier.

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Belgium, Canada, France, Japan, Monaco, Netherlands, Qatar, United Arab Emirates, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2024-08-30
Report date
2024-10-23
Voluntary/Mandated
Voluntary: Firm initiated
Location
South Jordan, UT, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0058-2025