FDA recall Z-0062-2019

Dannoritzer Medizintechnik GmbH & Co. KG · Class III · device

Product

Laparoscopy Instruments Instructions For Use (IFU) for the following: 1)Monopolar electrode, 2) Insert, handle and/or shaft component of a modular forceps or scissors for electrosurgical coagulation, 3) One-piece forceps or scissors for electrosurgical coagulation

Reason for recall

The Manual cleaning process and the usability of the reprocessing instruction by the user had not been validated. The IFU was modified to remove the manual cleaning method and revised the instructions for use for clarity and readability.

Distribution

US Nationwide in the States of OH, MI, MA, CA, OK, TN, GA, FL, IL, NC, TX, MO.

Key facts

Status
Terminated
Initiation date
2018-08-06
Report date
2018-10-10
Termination date
2019-07-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tuttlingen, Germany

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0062-2019