FDA recall Z-0066-2019

Johnson & Johnson Surgical Vision Inc · Class II · device

Product

OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215; 44107315; and 44065613 label states "Manufactured for: OptiMedica Corporation 1310 Moffett Park Drive Sunnyvale, CA 94089 USA Product Label for Serial Numbers: 46120916; 46123116, 46124617; 46131418, and 46131618 label states "Manufactured for: OptiMedica Corporation 510 Cottonwood Drive Milpitas, CA 95035 USA The Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.

Reason for recall

Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.

Distribution

US Distribution to states of: CO, GA, IL, MD including Puerto Rico and internationally to countries of: Austria and Germany.

Key facts

Status
Terminated
Initiation date
2018-05-09
Report date
2018-10-10
Termination date
2021-01-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Milpitas, CA, United States

Primary source