FDA recall Z-0067-2019

Heartware, Inc. · Class II · device

Product

Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.

Reason for recall

Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.

Distribution

Worldwide Distribution

Key facts

Status
Terminated
Initiation date
2018-08-28
Report date
2018-10-10
Termination date
2019-04-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Miami Lakes, FL, United States