FDA recall Z-0075-2021

Biomet 3i, LLC · Class II · device

Product

CERTAIN BELLATEK TIN ABUTMENT 4.1MM Reference Number: IEDAN4

Reason for recall

Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient

Distribution

FL, GA, ID, IL, MI, NV, NY, OH, SC, TN, VA

Key facts

Status
Terminated
Initiation date
2020-08-31
Report date
2020-10-14
Termination date
2022-07-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Palm Beach Gardens, FL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0075-2021