# FDA recall Z-0085-2021

> **Roche Diagnostics Operations, Inc.** · Class II · device recall initiated 2020-08-31.

## Product

BenchMark ULTRA Advanced Staining System (IVD)-  Material Number: 05342716001

## Reason for recall

Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket

## Distribution

Nationwide  Foreign:  Canada,   Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK,  Algeria, Argentina, Aruba, Australia, Bahrain, Bosnia-Herz., Brazil, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Hong Kong, India, Israel, Japan, Jordan, Kazakhstan, Kuwait, Malaysia, Martinque, Mexico, Monaco, Morocco, Namibia, New Caledonia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Russia, Saudi Arabia, Singapore, S. Africa, S. Korea, Taiwan, Thailand, Ukraine, UAE, Vietnam

## Key facts

- **Recall number:** Z-0085-2021
- **Recalling firm:** Roche Diagnostics Operations, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2020-08-31
- **Report date:** 2020-10-14

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Indianapolis, IN, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0085-2021

## Citation

> AI Analytics. FDA recall Z-0085-2021. Retrieved 2026-07-20 from https://api.ai-analytics.org/recall/Z-0085-2021. Source: US FDA. Licensed CC0.

---