FDA recall Z-0087-2023

Shanghai United Imaging Healthcare Co., Ltd. · Class II · device

Product

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Reason for recall

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

Distribution

Distribution to US states of TX, WI, CA, and MI

Key facts

Status
Ongoing
Initiation date
2022-05-30
Report date
2022-10-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Shanghai, N/A, China