FDA recall Z-0089-2023

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A

Reason for recall

Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.

Distribution

Nationwide Foreign: Canada

Key facts

Status
Ongoing
Initiation date
2022-08-24
Report date
2022-10-26
Voluntary/Mandated
Voluntary: Firm initiated
Location
Newark, DE, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0089-2023