FDA recall Z-0093-2024

CareFusion 303, Inc. · Class II · device

Product

BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01; Enterprise Server, REF: 1115-00, 134157-01, 137670-02, 134148-01, 135292-01, 136449-01, 133760-01

Reason for recall

When global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty), 4) CII Safe Stock Out Notice are, without user warning, overwritten to the value NO , which could result in diversion of controlled substances from the automated dispensing cabinet.

Distribution

US Nationwide distribution in the states of LA, NC, NY, MI, MA, NV, KS, IA, CA.

Key facts

Status
Ongoing
Initiation date
2023-09-06
Report date
2023-10-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States