FDA recall Z-0097-2021

Biomet, Inc. · Class II · device

Product

(1) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Bi-Polar Cup, E1 Antioxidant Infused, 28 MM, 41 MM: Item No. 110010458; Lot No. 710930; UDI No. (01) 00880304568747 (17) 230614 (10) 710930 (2) RingLoc Hip system Acetabular Bi-Polar Cup: Acetabular Cup, ArCom, 41 MM OD, 28 MM ID: Item No. 11-165206, Lot No. 649750; UDI No. (01) 00880304001923 (17) 230614 (10) 649750 (3) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Cup, ArCom, 52 MM OD, 28 MM ID: Item No. 11-165228; Lot No. 433290; UDI No. (01) 00880304001930 (17) 230612 (10) 433290

Reason for recall

Zimmer Biomet is conducting a medical device recall for three lots of the RingLoc Bi-Polar Hip system Acetabular Cup. The affected lots were gamma sterilized twice. Sufficient data does not exist to support the functionality, shelf life, or package integrity for more than one-time sterilization.

Distribution

Devices were distributed nationwide throughout the United States and to locations OUS. There has been one unit returned prior to the recall and approximately 14 units that have been previously implanted, leaving 33 units in the field subject to removal.

Key facts

Status
Terminated
Initiation date
2020-09-03
Report date
2020-10-21
Termination date
2022-12-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States