FDA recall Z-0102-2026

Boston Scientific Corporation · Class I · device

Product

Model Number U128, VALITUDE CRT-P EL MRI

Reason for recall

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Distribution

Worldwide

Key facts

Status
Ongoing
Initiation date
2025-08-20
Report date
2025-10-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Paul, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0102-2026