FDA recall Z-0103-2021

Roche Diagnostics Operations, Inc. · Class III · device

Product

Tina-quant Complement C4 ver.2, Catalog 05991994190

Reason for recall

Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Distribution

Domestic distribution nationwide.

Key facts

Status
Terminated
Initiation date
2020-08-27
Report date
2020-10-21
Termination date
2024-05-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Indianapolis, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0103-2021