FDA recall Z-0109-2022

Draeger Medical, Inc. · Class II · device

Product

Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400

Reason for recall

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Distribution

US Nationwide

Key facts

Status
Ongoing
Initiation date
2021-09-10
Report date
2021-10-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Telford, PA, United States

Primary source