FDA recall Z-0114-2018

3M Company - Health Care Business · Class II · device

Product

3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)

Reason for recall

During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.

Distribution

CA, MI, IL, MO, MN, GA, and Canada, Switzerland

Key facts

Status
Terminated
Initiation date
2017-10-17
Report date
2017-11-29
Termination date
2020-11-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Paul, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0114-2018