FDA recall Z-0114-2020

GE Healthcare, LLC · Class II · device

Product

Aestiva MRI, Model Numbers: a) 1006-9310-000 b) 1006-9110-000 c) 1006-9023-000 d) 1006-9028-000 e) 1006-9310-000-305077 f) 1006-9310-000-015243 g) 1006-9310-000-017602 h) 1006-9310-000-103785 I) 1006-9310-000-025109 j) 1006-9310-000-009650 k) 1006-9310-000-015224 l) 1006-9310-000-031881 m) 1006-9310-000-031854 n) 1006-9310-000-026571

Reason for recall

Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.

Distribution

Worldwide

Key facts

Status
Terminated
Initiation date
2019-07-19
Report date
2019-12-25
Termination date
2026-04-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source