FDA recall Z-0115-2021

Siemens Healthcare Diagnostics, Inc · Class II · device

Product

Atellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes: (01)00630414599151(10)98034139(17)20210130 and (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10995592, UDI Code (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10308994, UDI codes : (01)00630414471235(10)43157138(17)20210130 (01)00630414471235(10)81653138(17)20210130 (01)00630414471235(10)99728138(17)20210130 (01)00630414471235(10)24043138(17)20210130 (01)00630414471235(10)32477138(17)20210130 Advia Centaur Systems HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10308993, UDI Codes: (01)00630414471228(10)64661A48(17)20201112 (01)00630414471228(10)81211A48(17)20201112 (01)00630414471228(10)91743A48(17)20201112 (01)00630414471228(10)22124A49(17)20201211 Product Usage: This product is intended to be used for in vitro diagnostic use in calibrating the ADVIA

Reason for recall

Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results.

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2020-08-25
Report date
2020-10-28
Termination date
2023-03-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
East Walpole, MA, United States