FDA recall Z-0131-2023

Datascope Corporation · Class II · device

Product

LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter.

Reason for recall

The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy. Therapy delays may impact those patients who are more hemodynamically unstable.

Distribution

No US distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Columbia, Ecuador, Egypt, France, Indonesia, Israel, Italy, Jordan, Kenya, Malaysia, Mexico, Netherlands, Pakistan, Poland, Portugal, South Korea, Russia, Spain, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay.

Key facts

Status
Ongoing
Initiation date
2022-08-05
Report date
2022-11-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Fairfield, NJ, United States