FDA recall Z-0132-2026

Insulet Corporation · Class II · device

Product

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

Reason for recall

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2025-09-11
Report date
2025-10-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Acton, MA, United States

Primary source