FDA recall Z-0140-2018

Exactech, Inc. · Class II · device

Product

Optetrak Offset Tibial Tray and Screws, Size 4F/4T, Alpha, Ref 208-04-43, Sterile, RX.

Reason for recall

There is a design incompatibility between certain sizes of the offset spine stiffener screw and the constrained condylar tibial insert when used in conjunction with an offset tibial tray.

Distribution

Distribution was made to AL, FL, IN, ME, MN, NJ, NY, OH, OK, VA, and Puerto Rico. Foreign distribution was made to Argentina, Australia, Brazil, China, Colombia, France, India, Italy, Luxembourg, The Netherlands, Spain, Switzerland, Tunisia, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2017-09-19
Report date
2017-11-29
Termination date
2021-06-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Gainesville, FL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0140-2018