FDA recall Z-0154-2023

Insulet Corporation · Class I · device

Product

18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.

Reason for recall

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Distribution

Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.

Key facts

Status
Terminated
Initiation date
2022-10-17
Report date
2022-11-23
Termination date
2024-09-17
Voluntary/Mandated
Voluntary: Firm initiated