FDA recall Z-0169-2023

Johnson & Johnson Surgical Vision, Inc. · Class II · device

Product

THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in conjunction with an RF generator, for cardiac ablation.

Reason for recall

Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.

Distribution

Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY.

Key facts

Status
Ongoing
Initiation date
2022-09-07
Report date
2022-11-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irvine, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0169-2023