FDA recall Z-0171-2020

ICU Medical Inc · Class II · device

Product

Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.

Reason for recall

A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.

Distribution

US nationwide distribution, and Canada

Key facts

Status
Terminated
Initiation date
2018-08-01
Report date
2019-10-30
Termination date
2021-04-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lake Forest, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0171-2020