# FDA recall Z-0181-2023

> **Fresenius Medical Care Holdings, Inc.** · Class II · device recall initiated 2022-10-04.

## Product

NaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case.  For dialysis.

## Reason for recall

The "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.

## Distribution

Domestic distribution to AL AZ CA DC FL GA HI IL IN MD MI MO MS NE NJ NY OH PA SD TX VA  No foreign distribution.

## Key facts

- **Recall number:** Z-0181-2023
- **Recalling firm:** Fresenius Medical Care Holdings, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2022-10-04
- **Report date:** 2022-11-16

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Waltham, MA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0181-2023

## Citation

> AI Analytics. FDA recall Z-0181-2023. Retrieved 2026-06-26 from https://api.ai-analytics.org/recall/Z-0181-2023. Source: US FDA. Licensed CC0.

---