FDA recall Z-0182-2022

Elekta, Inc. · Class II · device

Product

Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00

Reason for recall

Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.

Distribution

Distribution throughout USA OUS distribution including:Algeria, Argentina, Australia, Austria Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam

Key facts

Status
Ongoing
Initiation date
2021-10-13
Report date
2021-11-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Atlanta, GA, United States