FDA recall Z-0185-2022

Cook Inc. · Class II · device

Product

Flexor Check-Flo Introducer, Introducer Set, RPN KCFW-6.0-35-45-RB, GPN G09908 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator.

Reason for recall

Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.

Distribution

Domestic distribution to AR AZ CT FL GA IL LA MA ME MN ND NJ NY OH PA TX WA. Foreign distribution to Germany, Finland, Italy, Netherlands, and Switzerland.

Key facts

Status
Terminated
Initiation date
2021-09-28
Report date
2021-11-10
Termination date
2024-02-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bloomington, IN, United States